DT5560 Update Packet
Software update for a DT5560 desktop digitizer (Zynq 7030) that is already installed and running. The packet refreshes the display and system manager application, the USB network setup, the x5560 web dashboard, the unified nibridge SDK bridge, Resource Explorer and the default MCA firmware. It is uploaded from the web dashboard of the DT5560, manually or with the online update (FOTA), and is applied by the instrument at the next reboot.
Compatibility
| Product | Board | Variant | Device | Operating system |
|---|---|---|---|---|
| DT5560 (DT5560SE on this portal) | DT5560 board | single | Zynq 7030 | PetaLinux 2019.2 DT5560 image (dt5560-image) |
This packet is for the DT5560 only; R5560 and R5560SE boards have their own update packets. It
does not update the Intel MAX10 FPGA of the DT5560 base board: that firmware has its own field
update, DT5560_Update_Base_Fw.RPD in the package dt5560-max10-base-firmware.
The dashboard of the DT5560 reports the model as DT5560SE on the Device Status page.
What's in this package
| Artifact | File | OS / architecture |
|---|---|---|
dt5560-update-packet | DT5560_Update.NIU | linux-zynq / armhf |
Despite the .NIU extension, DT5560_Update.NIU is not an FPGA firmware package: it is a
gzip-compressed tar archive (legacy update format, without a signed header) that contains an
update/ folder and the installation script run_update.sh. At the next boot the instrument
installs the following files on its data partition (/mnt/sdcard):
| Content | Installed to | Purpose |
|---|---|---|
nidisp.elf | /mnt/sdcard/software/ni/ | display and system manager application (touch display, analog settings, firmware loading, built-in web server) |
usb.sh | /mnt/sdcard/software/ni/ | USB network setup: RNDIS Ethernet gadget on the USB port, with a DHCP server for the host computer |
MCA.niu | /mnt/sdcard/user_firmware/MCA.niu | default MCA FPGA firmware of the DT5560 (dt5560-mca-firmware) |
resource_explorer/ (resource-explorer.elf, sw.json, libr5560.so, www/) | /mnt/sdcard/software/ni_app/resource_explorer/ | Resource Explorer server, its web interface and the libr5560 library |
ni-dashboard, ni-dashboard-loop.sh, config.example.json, VERSION, keys/ | /mnt/sdcard/software/ni/ | x5560 web dashboard server (with its supervision script, example configuration, version file and key folder) |
www/ | /mnt/sdcard/software/ni/www/ | x5560 web dashboard user interface |
nibridge (unified) | /mnt/sdcard/software/ni/nibridge | unified SDK bridge; installed only when the operating system starts the dashboard (see below) |
nibridge.conf (dt5560.conf profile) | /mnt/sdcard/software/config/nibridge.conf | DT5560 profile of the unified nibridge; same condition |
legacy_web.json | /mnt/sdcard/software/config/legacy_web.json | moves the built-in web server of the display application to 127.0.0.1:8080, behind the dashboard; same condition |
nibridge (DT5560 bridge) | /mnt/sdcard/software/ni/nibridge | previous DT5560 SDK bridge; installed instead of the unified one on an older operating system |
After copying, the script sets the permissions of the data partition, then deletes the update/
folder and itself.
How to install
Before you start
- Download
DT5560_Update.NIUfrom NI Packages (only for the manual procedure; FOTA downloads it for you). - Plan a reboot of the DT5560: running acquisitions are interrupted.
Manual upload from the web dashboard
- Open the dashboard of the DT5560 in a browser:
http://<dt5560-ip>/orhttps://<dt5560-ip>/. - Log in. At the first login with the factory default password the dashboard asks you to set a new password.
- Go to Device Update (not Firmware: the Firmware page accepts only FPGA firmware
packages and refuses this file). In Manual update click Choose an update file or drop it
here and select
DT5560_Update.NIU. - The instrument reads the beginning of the file and shows Type Legacy update (unsigned) and Format legacy (no header), with the warning to install it only if it comes from a trusted source. Packets downloaded from NI Packages are such a source.
- Click Upload and stage and confirm. Do not close the page during the transfer.
- The page shows Update uploaded and verified. and the packet under Staged update, with the note "The update is applied at the next reboot. Do not turn off the device until the procedure ends." An administrator can still remove it with Discard staged update.
- Reboot the DT5560: Reboot button (administrator) or a power cycle. Do not switch the instrument off while it is starting up after the update.
Online update (FOTA)
The browser, not the instrument, contacts the update server https://downloads.nuclearinstruments.eu:
the DT5560 does not need Internet access, the computer running the browser does.
- Open the dashboard and go to Device Update (steps 1 to 3 above).
- In Online update (FOTA) click Check for updates. The card shows the available release with its name, version and Changelog.
- Click Download and install. The browser downloads the packet, verifies its SHA-256 and uploads it to the instrument. A checksum error stops the procedure and nothing is staged.
- Reboot the DT5560 as in step 7 above.
Check the result
- After the reboot open Device Update: Last result shows the Time and the Result (success or failed, with the Error text). The Version field stays empty for this packet format.
- On Device Status check the Dashboard version.
- On Firmware the file
MCA.niuis listed. If another firmware is active and you want the default MCA firmware, selectMCA.niu, click Install and reboot the instrument.
What is preserved
The installation script only copies the files listed above. Everything else on the data partition
is kept: dashboard settings and accounts, network and analog settings, other firmware files in the
firmware folder and user data. Files with the same name are replaced: in particular MCA.niu and
the nibridge configuration /mnt/sdcard/software/config/nibridge.conf, so local changes to that
file must be applied again after the update.
The packet does not change the operating system, the boot files, the ADC initialisation program, the display logo or the MAX10 base firmware.
Instruments with an older operating system
On instruments whose operating system does not yet start the x5560 dashboard, the packet installs
it but the existing web interface stays in use: the built-in web server keeps port 80 and the
previous DT5560 SDK bridge is installed instead of the unified nibridge. The full dashboard becomes
active with an operating system image that includes it (dt5560-image). On such instruments upload
the packet with the update function of the existing web interface.
Requirements
- A DT5560 running its PetaLinux 2019.2 operating system.
- A computer whose browser reaches the DT5560 over the network; for FOTA the same computer must
also reach
https://downloads.nuclearinstruments.eu. - A dashboard account: any logged-in user can upload and stage the packet; rebooting from the
dashboard and discarding a staged update require the administrator account (
root). - The security option Require signed updates (Security page) must be off: when it is on, unsigned and legacy packets such as this one are refused.
- With HTTPS the browser shows a certificate warning for the self-signed certificate of the instrument unless a custom certificate is installed on the Security page; plain HTTP also works.
Versioning and releases
- Releases are tagged with the date format
YYYY.MM.DD.NN, for example2026.10.11.00. - The changelog (
CHANGELOG.md) is shown on NI Packages with each release. - Release channel on this portal: stable.
- The packet does not carry a version header, so Last result shows no version, and the packet does not change the Software version shown on Device Status. The Dashboard version on Device Status is the version of the dashboard installed by the packet.
Related packages
| Package | Relation |
|---|---|
dt5560-image | production image of the DT5560; its data partition overlay provides usb.sh |
dt5560-linux-apps | display and system manager application and previous SDK bridge, included |
dt5560-mca-firmware | default MCA firmware, included as MCA.niu |
dt5560-max10-base-firmware | MAX10 base board firmware and its separate field update packet |
x5560-dashboard-backend | dashboard server, included |
x5560-dashboard-frontend | dashboard user interface, included |
nibridge | unified SDK bridge, included with the dt5560.conf profile |
libr5560 | the board-side libr5560.so used by Resource Explorer is included |
resource-explorer-server, resource-explorer-web | Resource Explorer server and web interface, included |
Known limitations and notes
- The packet uses the legacy unsigned format: the instrument cannot verify product and signature
before installation. Install only
DT5560_Update.NIUfiles obtained from NI Packages. - The packet is refused while Require signed updates is enabled.
- The update does not refresh the factory copy of the data partition kept on the boot partition. If the instrument ever has to rebuild its data partition from that copy, it returns to the software of its production image: install the packet again.
- Because the Software version shown by the instrument does not change with this packet, the online update may keep proposing the latest release after it has been installed; use Last result to confirm the installation.
Source and support
- Development repository: https://gitlab.lanni/nuclear-instruments/prodotti/x5560/fw-sw/dt5560/dt5560-update-packet
- For support contact Nuclear Instruments / CAEN support.
DT5560 Update Packet
Update packet for a DT5560 already in the field: applications, USB setup, web app, Resource Explorer and default MCA firmware. Uploaded from the dashboard and applied at the next boot.
License:
Trending Products
DAQ14125
High-performance 14-bit, 125 MS/s digitizer with 8 single-ended analog input channels, designed for precision data acquisition and customizable real-time signal processing.
DAQ121
High-performance, fully integrated digitizer developed by Nuclear Instruments for advanced SiPM-based muon detection systems.
Sci-Compiler
FPGA firmware generation tool for scientific instrumentation and data acquisition.
Trending Packages
Lib R5560
Cross-platform hardware access library for communicating with CAEN R5560 and DT5560 devices.
Quanta
Acquisition and analysis software for the Nuclear Instruments digitizer platform — the software layer of the QuantumBox chain.
Scintilla
Scintilla: desktop control center for one or more I-Spector spectrometers, over USB or Ethernet.